Barcode Scanner Information
Reagent Bay and Sample Bay Barcode Scanner Information
All Sample Bay and Reagent Bay bar code scanners are to be replaced on Failure ONLY.
All scanners removed from Instruments must be returned immediately to the Hologic Dx Repair Center to be repaired.
|
ALL Systems Serial #01854 and above require a THIN Sample Bay Bar Code Scanner |
|
All Panther Fusion Systems require a THIN Sample Bay Bar Code Scanner |
|
THIN Sample Bay Bar Code Scanners are pre-installed (during manufacture) on all Systems Serial #02590 and above. |
Below is a summary of the scanners we have released:
| Holx PN | Rev | BCL # | Hologic Description |
|---|---|---|---|
| 902936 / R902936 | 1 | BCL148.60 | Sample/Reagent Bay, Bar Code Scanner |
| MME-02832 / RMME-02832 |
1 | BCL148.70 | Sample/Reagent Bay, Bar Code Scanner |
| MME-03470 / RMME-03470 |
1 | BCL148.71 | Reagent/Sample Bay, Bar Code Scanner |
| MME-03470 / RMME-03470 |
2 | BCL148.72, v81.18 | Reagent/Sample Bay, Bar Code Scanner |
| ASY-09778 / RASY-09778 |
1 | BCL148.75 | Fusion Sample Bay, Bar Code Scanner |
| MME-03417 | 1 | BCL148.76 | Sample Bay, Bar Code Scanner |
| MME-03472 / RMME-03472 |
1 | BCL148.76 | Sample Bay, Bar Code Scanner |
| MME-03472 / RMME-03472 |
2 | BCL148.7, v81.18 | Sample Bay, Bar Code Scanner |
| MME-04064 | 1 | BCL148.7, v82.18 | Sample/Reagent Bay, Barcode Scanner |
| RMME-04064 | BCL148.71, V82.18 BCL148.72, V82.18 | Sample/Reagent Bay, Barcode Scanner | |
| MME-04065 | BCL148.78, V82.18 | Sample/Reagent Bay, Barcode Scanner THIN | |
| RMME-04065 | BCL148.76, V82.18 BCL148.77, V82.18 | Sample/Reagent Bay, Barcode Scanner THIN |
All version of scanners should be returned as RMA.
Below is an example of how the Barcode Scanners will be labeled.